Home
International Journal of Science and Research Archive
International, Peer reviewed, Open access Journal ISSN Approved Journal No. 2582-8185

Main navigation

  • Home
    • Journal Information
    • Abstracting and Indexing
    • Editorial Board Members
    • Reviewer Panel
    • Journal Policies
    • IJSRA CrossMark Policy
    • Publication Ethics
    • Instructions for Authors
    • Article processing fee
    • Track Manuscript Status
    • Get Publication Certificate
    • Current Issue
    • Issue in Progress
    • Past Issues
    • Become a Reviewer panel member
    • Join as Editorial Board Member
  • Contact us
  • Downloads

ISSN Approved Journal || eISSN: 2582-8185 || CODEN: IJSRO2 || Impact Factor 8.2 || Google Scholar and CrossRef Indexed

Fast Publication within 48 hours || Low Article Processing Charges || Peer Reviewed and Referred Journal || Free Certificate

Research and review articles are invited for publication in January 2026 (Volume 18, Issue 1)

Radiological safety assessment for safe transportation of Co-60 disused sealed radioactive sources of teletherapy machines

Breadcrumb

  • Home
  • Radiological safety assessment for safe transportation of Co-60 disused sealed radioactive sources of teletherapy machines

M. Shamsuzzaman 1, *, M. H. Rahman 1, B. C. Sutradhar 1, K. Assaduzzaman 1, M. M. H. Bhuiyan 1 S. M. Shome 1, M. M. Islam 2 and D. Paul 3

1 Health Physics and Radioactive Waste Management Unit, Institute of Nuclear Science and Technology, AERE

2 Planning Division, Bangladesh Atomic Energy Commission, Dhaka-1207

3 Atomic Energy Research Establishment (AERE), Savar, Dhaka-1349

Research Article

International Journal of Science and Research Archive, 2025, 15(01), 323-332

Article DOI: 10.30574/ijsra.2025.15.1.0006

DOI url: https://doi.org/10.30574/ijsra.2025.15.1.0006

Received on 30 January 2025; revised on 07 March 2025; accepted on 10 March 2025

The radiological circumstances of Co-60 teletherapy source heads were inspected to ensure their safe transportation from four medical college hospitals to CWPSF. In this perspective, shielding containers for four Co-60 source heads were manufactured, and inspected their radiological safety in terms of dose equivalent rates (DERs), as well as Transport Index (TI). In case of DMCH, CMCH, SMAGOMCH and RMCH, maximum DER on container surface were found to be 10.00 µSvh-1, 13.30 µSvh-1, 17.16 µSvh-1, 25.70 µSvh-1 with observed TI 0.03, 0.05, 0.06, 0.08, respectively. Transport safety was verified by measuring DER on the vehicle surface, as observed in the maximum range of 0.11 - 3.13 µSvh-1  ; and 0.08 – 0.25 µSvh-1 at 2 m distance for RMCH, which is significantly less than the IAEA permissible limit [12]. Consequently, DER was further verified at driver’s position, as observed 0.11 – 0.17 µSvh-1 for head, 0.14- 0.16 µSvh-1for Chest, 0.11 - 0.13 µSvh-1 for gonad, and 0.09 – 0.11 for Leg, which are remarkably lesser than the IAEA and ICRP limit of 10 µSvh-1 and 0.5 µSvh-1 for occupational and public, respectively. Therefore, the investigated DERs indicated no significant radiological safety concern as per international dose limits [6, 7], and IAEA guidelines for safe transportation of radioactive materials [12].

Radiological safety; Cobalt-60 teletherapy source; Gamma dose equivalent rate; Transport index; Ionizing radiation

https://journalijsra.com/sites/default/files/fulltext_pdf/IJSRA-2025-0006.pdf

Preview Article PDF

M. Shamsuzzaman, M. H. Rahman, B. C. Sutradhar, K. Assaduzzaman, M. M. H. Bhuiyan, S. M. Shome, M. M. Islam and D. Paul. Radiological safety assessment for safe transportation of Co-60 disused sealed radioactive sources of teletherapy machines. International Journal of Science and Research Archive, 2025, 15(01), 323-332. Article DOI: https://doi.org/10.30574/ijsra.2025.15.1.0006.

Copyright © 2025 Author(s) retain the copyright of this article. This article is published under the terms of the Creative Commons Attribution Liscense 4.0

For Authors: Fast Publication of Research and Review Papers


ISSN Approved Journal publication within 48 hrs in minimum fees USD 35, Impact Factor 8.2


 Submit Paper Online     Google Scholar Indexing Peer Review Process

Footer menu

  • Contact

Copyright © 2026 International Journal of Science and Research Archive - All rights reserved

Developed & Designed by VS Infosolution